Product description:
I. overview
UDZ-10-21/ type electric contact double color level meter, mainly used for a variety of high and low water level, heater, deaerator, evaporator and row water tank water, liquid level measurement, compared with similar products and has the advantages of simple operation, long service life, high safety factor.
Two, structure
The water level gauge is mainly composed of a measuring cylinder, ceramic electrode, steam, water blue, two instruments etc..
Three, the principle structure
Structural schematic diagram of measuring cylinder
Four, working principle
The instrument by using the characteristic conductivity difference of boiler water and steam are measured, due to the change of liquid level to the electrode immersed in water, part of the electrode placed in steam, boiler water electrodes immersed in the cylinder and the impedance decreases, the impedance of the electrode in the steam cylinder. Using this feature, the non electric water level can be converted into electricity, intelligent two meters, so as to achieve the water level display, alarm, protection, chain function.
The use of single-chip microcomputer control, through the use of the hardware watchdog can be automatically reset when running in the microcontroller program to ensure that the instrument working condition is more secure and reliable. At the same time, the thermal steam red, water used to mark the double color synthesis column, visually display the steam and water level.
Five, characteristics
1 the use of red, green double color synthesis column display level, red digital display level coordinates, solves the operator to monitor the running state of eye fatigue, intuitive and clear visual effect.
2 has the alarm alarm (sound) (relay output), silencing, self-test, screening function. Compared with the old two meters, the utility model has the advantages that the operation is more convenient and intuitive, and the alarm value can be set more than once.
3 based on the characteristic of CMOS high input impedance, current signal input circuit only weak through the electrode, can reduce the chemical corrosion electrode to a minimum, therefore, the two meter could more significantly prolong the service life of the electrodes.
4 by fuzzy control, according to the electricity and water resistance threshold of different water quality automatic electrode regulator, to the needs of different applications.
Six, technical parameters
Measuring tube technical parameters
Product type: UDZ-10-21/
Working pressure: ≥ 2.5Mpa≤ 32Mpa
Tube specification: Φ 89× 12; Φ 108×
Connecting pipe specification: Φ 38× 5
Blowdown pipe specification: Φ 28× 4 20G
Electrode installation: divided into two types
Two meter technical parameters
Working environment:
Ambient temperature: -10 C ~ +45 C
Relative humidity: 80%.
Power supply: 220VAC± 10%, 50Hz current 1A
Power consumption: 10VA
Measured liquid impedance conditions: 0 ~ 90KΩ
Installation: vertical disc
Dimensions: 280*80*160 (mm) (L * w * h)
Instrument panel opening size: 152*76 (mm)